Diagnostic accuracy of pre-hospital invasive arterial blood pressure monitoring for haemodynamic management in traumatic brain injury and spontaneous intracranial haemorrhage
Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine volume 33, Article number: 89 (2025)
Abstract
Background
Neuroprotective measures to prevent secondary brain injury are a critical aspect of pre-hospital management in patients with acute traumatic brain injury (TBI) and spontaneous intracranial haemorrhage (sICH). Haemodynamic optimisation guided by non-invasive blood pressure (NIBP) measurements is an important neuroprotective measure, as cerebral autoregulation is often absent or impaired. The accuracy and clinical relevance of invasive arterial blood pressure (IBP) monitoring to optimise haemodynamic management has not been established in patients with a brain insult.
Methods
A retrospective clinical diagnostic accuracy study to establish the accuracy and clinical relevance of IBP-guided haemodynamic optimisation in patients with TBI or sICH. The occurrence- and clinical relevance of IBP-NIBP discrepancies in patients attended by a UK Helicopter Emergency Medical Service (HEMS) between 6 January 2022 and 6 January 2024 was evaluated. Bland-Altman plots with adjustment for repeated measures were constructed to analyse disagreement in relation to absolute blood pressure values. Multivariate analysis was performed using generalised linear mixed effects regression (GLMER) models with random effects to identify predictors of disagreement. Error Grid Analysis (EGA) classified the clinical relevance of discrepancies. The primary outcome was pairwise agreement between IBP and NIBP, defined as less than 10% difference in mean arterial pressure (MAP).
Results
For 209 patients (159 TBI and 50 sICH) 1020 concurrent IBP and NIBP measurements were available. The average [95% CI] difference in MAP was -1.4 mmHg (-3.09 to 0.27) and 2.6mmHg in TBI. Only 459 (54.7%) MAP data met criteria for pairwise agreement. Multivariate regression analysis revealed a strong association between MAP disagreement and ground emergency medical service conveyance (aOR 2.01, 95% CI 0.98-4.10). Bland-Altman analysis demonstrated proportional bias, with NIBP underestimation of MAP at higher blood pressures and overestimation at lower blood pressures. EGA revealed that in 6.1% (95% CI: 4.5-7.7) of TBI and 12.5% (95% CI: 7.8-17.2) of patients with sICH pairwise disagreement was associated with a moderate to dangerous risk of over- or undertreatment.
Conclusion
NIBP guided pre-hospital haemodynamic management of patients with TBI or sICH is hampered by clinically relevant measurement inaccuracies in a significant proportion of patients. Pre-hospital IBP has the potential to improve early haemodynamic optimisation, especially when hypo- or hypertension is present, enabling tailored neuroprotection in the hyperacute phase.
Background
Neuroprotective measures to prevent secondary brain injury are a critical aspect of pre-hospital management in patients with acute traumatic brain injury (TBI) and spontaneous intracranial haemorrhage (sICH) [1, 2]. Haemodynamic optimisation is an important neuroprotective measure in TBI patients. Cerebral autoregulation is absent or impaired in 49–87% of these patients [1], therefore maintaining appropriate mean arterial pressure (MAP) is crucial to preserve adequate cerebral blood flow (CBF) and prevent secondary brain injury [3,4,5].
Pre-hospital critical care teams such as helicopter emergency medical services (HEMS) have the capability to provide both volume replacement and vasopressor therapy guided by non-invasive blood pressure measurements (NIBP) [2]. However, NIBP measurements can be confounded by movement artefact, and vibratory effects during transport, and may be less reliable in patients with severe hypo and hypertension [6,7,8,9,10]. This may result in artefactual or intermittent measurements which either over- or underestimate blood pressure values [3,4,5], with important consequences for subsequent initiation- or withholding of treatment [11] over extended conveyance intervals [6].
Invasive arterial blood pressure (IBP) monitoring may offer a more reliable alternative [5, 7]. Retrospective studies have demonstrated the feasibility and efficiency of IBP in pre-hospital critical care [8,9,10]. Additionally, a recent study reported higher accuracy of IBP compared to NIBP in a heterogeneous cohort of pre-hospital patients [5]. However, two important questions remain unclear. First, the degree of inaccuracy in non-invasive measurements compared to invasive measurements. Second, whether such measurement differences translate to clinically relevant advantages for haemodynamic optimisation in patients with a brain insult.
This study aims to establish the accuracy and clinical relevance of IBP-guided haemodynamic optimisation in patients with TBI or sICH and determine if there is a perceived clinical advantage to guide neuroprotective measures in the pre-hospital phase.
Methods
Study setting and design
A retrospective observational diagnostic accuracy study was performed using data extracted from our pre-hospital electronic patient clinical record database (HEMSBase™ 3.0, MedicOne Systems). We analysed data from patients attended by HEMS in South-East England between January 6, 2022, and January 6, 2024. The data was collected as part of routine clinical care and retrospectively analysed for this study. Ambulance Charity Kent Surrey Sussex (KSS) provides pre-hospital advanced interventions within the associated major trauma systems in conjunction with ground emergency medical services (EMS). KSS operates across three counties serving a resident population of 5 million. A physician-paramedic pre-hospital emergency medicine team respond by either helicopter or rapid response vehicle. Doctors have a minimum of 5-years postgraduate experience, including a minimum of 6-months anaesthesia training. Paramedics undergo further specialist practice, including theoretical modules on invasive monitoring and pre-hospital emergency anaesthesia (PHEA). Subsequently, clinicians demonstrating clinical competency may then site pre-hospital invasive monitoring. Advanced interventions (including transfusion of blood components, airway management, anticoagulant reversal and surgical procedures) are delivered during the pre-hospital phase both at scene and during conveyance [12]. IBP monitoring was introduced in January 2022 in a governed stepwise manner following expert consensus by consultants and specialists in pre-hospital emergency medicine.
Patient selection
Patients were eligible for inclusion if they were 18 years or older and were attended by KSS between 6 January 2022 and 6 January 2024 for suspected TBI or sICH. Only patients in whom an arterial cannula was successfully sited and for whom concurrent IBP and NIBP measurements available were included. Paediatric patients and interhospital transfers were excluded from the present analysis [13].
Non-invasive and invasive arterial blood pressure monitoring
Non-invasive blood pressure measurements were performed using appropriately sized cuffs based on patient arm circumference, following manufacturer recommendations. The NIBP cuff is placed on the upper arm contralateral to the arterial line when possible. Measurements are taken at 3-min intervals unless manually triggered when clinically indicated.
According to KSS Standard Operating Procedure (SOP) invasive arterial blood pressure monitoring is indicated for patients with suspected TBI and sICH where HEMS clinicians perceive continuous invasive BP monitoring will contribute to haemodynamic optimisation. Arterial cannulation (Vygon SWITCH, 20 g, 45 mm length, Utah, US; Arterial Leadercath, 4 F, 10 cm Length, Vygon, France; or UltimumTM Hemostasis Introducer, 5 F, 12 cm length, Abbott Medical, USA) is commonly performed prior to PHEA, with the radial artery being the preferred access. The chosen device is connected to a disposable pressure monitoring set (TruWaveTM, PX260, Edwards Lifesciences, Irvine, CA, USA) via a 150-cm-long saline line with a pressure infusion bag (ACCU-PROTM, APPI00500, 500 ml, ProAct Medical Ltd, Northants, UK). The transducer is placed at the level of the right atrium. Both invasive- and non-invasive blood pressures are measured using the Tempus PROTM (Phillips Electronics, Farnborough, UK) device with arterial waveforms visible to clinicians in real-time. Data capture is at 3-min intervals for non-invasive and 1-min intervals for invasive monitoring.
Clinicians were trained to identify and address waveform artefacts, particularly issues related to overdamping and underdamping. Overdamping artefacts may include systolic blood pressure (SBP) underestimation, diastolic blood pressure (DBP) overestimation, slurred upstroke, absent dicrotic notch, and general loss of waveform detail. These were addressed by checking for underinflated pressure bags, air bubbles, blood clots, loose connections, and catheter kinking. Underdamping artefacts may include: SBP overestimation, DBP underestimation, exaggerated dicrotic notch, and non-physiological oscillations)Â and were addressed by examining tubing stiffness and transducer function.
Primary and secondary outcomes
The primary outcome was the percentage of concurrent blood pressure measurements demonstrating pairwise agreement between IBP and NIBP, defined as a <10 mmHg (MAP), or <20 mmHg (SBP and DBP) difference. Concurrent data pairs refer to one invasive and one non-invasive measurement recorded from the same patient within a 1-minute interval on the electronic patient clinical record (EPCR).
Secondary outcomes were:
-
1.
Pairwise agreement using alternative definitions (<20% or <30% difference for MAP and <30% or <40% difference for SBP/DBP).
-
2.
Patient- and/or treatment characteristics independently associated with pairwise disagreement.
-
3.
The presence- (and magnitude) of a relation between absolute blood pressure values and the magnitude of IBP-NIBP discrepancies (proportional bias).
-
4.
The estimated clinical relevance of any pairwise disagreement present, as quantified by Error Grid Analysis (EGA).
Data sources and data acquisition
Data was extracted through Zoho AnalyticsTM from the EPCR HEMSBaseTM 3.0 (MedicOne Systems) and included the following categories. Baseline demographic descriptors: age, sex and weight (clinician estimated). Physiological parameters included: Glasgow Coma Scale (GCS) score, Heart rate (HR), Oxygen saturation (SpO2), End-tidal Carbon dioxide (EtCO2), SBP, invasive SBP (iSBP), DBP, invasive DBP (iDBP), MAP and invasive MAP (iMAP). Pre-hospital advanced interventions: Arterial line catheterisation time and site [radial, brachial, femoral], type of arterial catheter used [Flowswitch, Leadercath], blood component transfusion [yes/no], vasoactive drug administration [yes/no].
Blood pressure measurements were screened for artefacts using a two-stage process. First, predefined thresholds were applied to exclude physiologically implausible measurements based on criteria from Juri et al. [14]: invasive SBP >300 mmHg or <20 mmHg, invasive DBP >225 mmHg or <5 mmHg, and pulse pressure (difference between iSBP and iDBP) <5 mmHg. These exclusion criteria were applied during data pre-processing. Second, the EPCR were reviewed by a panel of clinicians to identify artefactual invasive and non-invasive measurements, these pertained to measurements perceived to be physiologically implausible i.e. did not fit the physiological trend of the data.
Ethical considerations and data governance
Data were routinely collected and met Health Research Authority (HRA, UK) criteria for service evaluation. Research Ethics Committee approval was not required, and the project was approved by the KSS Research and Innovation Committee. Study conduct was in accordance with Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) [15].
Statistical analysis
Baseline characteristics (stratified for brain injury aetiology) were compared across groups using Chi Square (X2) test for categorical data and Mann-Whitney U test for continuous data. Mean differences (SD) between NIBP and IBP were calculated, and Bland-Altman analysis with adjustment for repeated measurements [16, 17] was performed to investigate the relation of mean differences with absolute blood pressure values. Mean differences were related to pre-defined limits of agreement between NIBP and IBP. Univariate regression analysis was performed using a generalised linear mixed model (GLMM) with binomial logit to identify variables related to disagreement at the pre-defined 10% acceptability threshold. Multivariate regression was performed with those factors that showed an R2 >0.20 in univariate analysis to investigate which variables were independent predictors of MAP disagreement.
Error grid analysis was performed to establish clinical relevance. Differences between invasive and non-invasive blood pressure measurements were classified into five risk levels, ranging from zone A (no risk, no difference in clinical action) to zone E (dangerous risk, unnecessary (absence of) treatment with severe life-threatening consequences) [18]. Risk zones were calibrated based on the aggregated results of a survey amongst 25 specialists in intensive care medicine and anaesthesiology) [18].
Data were pre-processed in Microsoft Excel (Microsoft Corp, Redmond, WA USA). Statistical analysis and visualisations were performed using R version 4.3 (R Core Team, 2023. R: A language and Environment for Statistical Computing. R Foundation for Statistical Computing, Vienna, Austria). R packages included: base r stats for logistic regression and model comparison, blandr for Bland-Altman Analysis, ggplot2 and flextable for creating plots and table creation, dplyr for data manipulation, broom for tidying model outputs, lme4 for mixed effects models. EGA was performed in Matlab 2019 a/b (The Mathworks Inc, Natwick, MA) [18]. Statistical significance was pre-determined at p <0.05.
Results
Baseline characteristics
During the study period, an arterial line was sited in 250/3840 patients attended by the service, 221 of which had a suspected TBI or sICH. A total of 209 were eligible for inclusion (Fig. 1). For these patients, a total of 1020 IBP-NIBP concurrent data pairs were available for comparison (median number of measurements per patient was 9 (IQR 3–15, range: 1–29).
Consort diagram for derivation of study population and frequency of concurrent data pairs. Legend. Derivation of study population and frequency of concurrent data pairs. IBP; invasive arterial blood pressure; EPCR, electronic patient clinical record. TBI; traumatic brain injury. sICH; spontaneous intracranial haemorrhage. iSBP; Invasive systolic blood pressure. iDBP; Invasive diastolic blood pressure
Baseline characteristics of the study population (stratified by aetiology of brain injury) are presented in Table 1. Most patients were male (185/209, 77%), with a median age of 51.2 [40–68] years. The radial artery was the preferred site for arterial catheterisation, and over 80% of the arterial lines were placed during the first attempt. In those patients whom received PHEA (198/209, 94.7%) an arterial line was sited pre-PHEA in 153/198 (77.3%). Vasopressors were used in 73 (44.5%) patients, and blood components were administered in 12 (5.8%) patients. Differences between invasive- and non-invasive average DBP, MAP and SBP are reported (Table 1). Average difference for TBI was 2.6 mmHg. Further, the mean absolute differences in MAP between non-invasive and IBP are represented in Table 2.
Using the most conservative acceptability threshold for pairwise agreement between IBP and NIBP, (<10% difference for MAP and <20% for SBP and DBP), 57.4% of the data pairs had a difference below the threshold for MAP, 77.0% for SBP and 65.5% for DBP. As expected, agreement increased with lower thresholds (higher acceptable disagreement), but even at a disagreement threshold of 30% for MAP, in 12.9% of all measurements the threshold was not met. SBP agreement was consistently higher then DBP agreement (Table 3).
Factors associated with pairwise disagreement
Univariate associations of patient variables with pairwise disagreement (MAP <10 mmHg) are presented in Additional file 1. In the multivariate mixed effects model only EMS conveyance of TBI patients was borderline associated with pairwise disagreement (OR 2.01, 95% CI 0.98–4.10, p = 0.05). In the sICH group, no fixed effects reached statistical significance (Fig. 2). The substantial random effects variance (1.61 for TBI and 1.92 for sICH) indicated considerable between-patient heterogeneity in MAP disagreement patterns. Forest plot of variables associated with pairwise disagreement in SBP and DBP are presented in Additional file 2.
Forest plot of variables potentially associated with pairwise disagreement in mean arterial pressure between invasive and non-invasive blood pressure in patients with suspected TBI and sICH (n=209). TBI, traumatic brain injury; sICH, spontaneous intracranial haemorrhage; CI, confidence interval; EMS, emergency medical service; PHEA, pre-hospital emergency anaesthesia. NB: Sex denotes female sex as reference value; Ground EMS denotes conveyance type with aircraft as reference value
Relation between absolute blood pressure and pairwise disagreement
To explore the relationship between absolute blood pressures and pairwise disagreement, Bland-Altman analysis with repeated measures was performed. In patients with TBI, the mean difference in MAP (accounting for the direction of the difference) between IBP and NIBP was 2.5 mmHg (limits of agreement −49.3 to 54.4 mmHg). A total of 66 concurrent measurements in 24 patients (6.4% of measurements, 15% of population) fell outside the limits of agreement. NIBP overestimated MAP in hypotensive patients, whereas it underestimated MAP in hypertensive patients (Fig. 3a). In patients with sICH, findings were similar (mean difference −3.5 mmHg, limits of agreement 56.4 to 63.4 mmHg) (Fig. 3b). Bland-Altman plots for SBP and DBP are shown in Additional file 3.
Bland-Altman plot of the difference between invasive and non-invasive mean arterial pressure measurements in patients with TBI (a) and sICH (b). Bland-Altman plots representing the mean difference in MAP (middle black horizontal line), the 95% limits of agreement (black solid lines) and bias (red dashed line) in TBI (a) and sICH (b)
Risk of pairwise disagreement
EGA for MAP in patients with TBI classified 69% (95% CI: 65.8–72.2), 24% (95% CI: 21.0–26.9), 4.3% (95% CI: 2.9–5.6), 1.5% (95% CI: 0.6–2.3), and 0.36% (95% CI: 0.0–0.7) of measurements in the risk categories A-E, respectively (Fig. 4). Based on this analysis, in 6.1% (95% CI: 4.4–7.7) of paired measurements NIBP differed from IBP with high clinical relevance (moderate to dangerous risk, categories C-E). For sICH, 12.5% (95% CI: 7.8–17.2) of paired measurements NIBP differed from IBP with high clinical relevance (moderate to dangerous risk, C-E). The cumulative percentage of MAP measurements falling within the low-risk categories (A-B) was 87.5% (95% CI: 82.8–92.2). EGA for SBP is available in Additional file 4.
Error grid analysis for mean arterial pressure comparing non-invasive and invasive measurements in patients with suspected TBI (a) and sICH (b). a Error Grid Analysis for TBI MAP (mmHg). b Error Grid Analysis for sICH MAP (mmHg). Invasive pressure (x-axis, reference standard) is plotted against non-invasive pressure (y-axis, index method). Data pairs marked by a black cross represent a concurrent data pair. Colour differentiation indicates the continuous risk level from the green (zone A, no risk) to dark red zone (zone E, dangerous risk). Zones A-E indicate: (A) no risk, no difference in clinical action, (B) low risk, benign or no treatment, (C) moderate risk, unnecessary treatment with moderate non-life-threatening consequences, (D) significant risk, unnecessary treatment with severe life-threatening consequences, (E) high risk, unnecessary treatment with severe life-threatening consequences. MAP; mean arterial pressure; TBI, traumatic brain injury; sICH, spontaneous intracranial haemorrhage
Discussion
Our findings suggest that when NIBP is used to guide pre-hospital haemodynamic optimisation in patients with TBI or sICH, clinically relevant measurement inaccuracies are observed in a significant proportion of patients, especially when hypo- or hypertension is present. To our knowledge, this is the first study to explore the use of pre-hospital IBP specifically in patients with a suspected brain insult. We suggest that pre-hospital invasive blood pressure monitoring has the potential to improve earlier haemodynamic optimisation, especially when hypo- or hypertension is present, further enabling tailored neuroprotection in the hyperacute phase.
As in previous (in-hospital) studies [18] the overall difference in average MAP between invasive- and non-invasive measurements in this study was small (-1.4 mmHg), and in accordance with thresholds set by the Association for the Advancement of Medical Instrumentation (AAMI) [19]. However, individual patient risk is not well expressed by this number, as the percentage of patients meeting criteria for pairwise disagreement between NIBP and IBP was substantial (42.6% of MAP values, 23% of SBP values and 34.4% of DBP values), reflecting a significant inter-individual variation, and potential for over- or undertreatment in a significant proportion of patients.
Pairwise disagreement was highest for MAP, and lowest for SBP. Differential reliability between blood pressure components is noteworthy given the measurement principles of modern pre-hospital monitors. These devices use oscillometric technology where MAP is directly measured at the point of maximum oscillations, while systolic and diastolic pressures are calculated using proprietary algorithms based on fixed ratios of oscillation amplitude [20]. MAP showed greater discrepancy, suggesting that oscillometric detection is vulnerable to motion artefacts during ambulance transport, while algorithmic SBP derivation appears more robust [20]. This is an important finding, as MAP is the main determinant of cerebral perfusion pressure (and thereby cerebral blood flow) [21, 22].
Of importance, pairwise agreement was related to absolute blood pressure values. Bland-Altman analysis demonstrated that NIBP overestimates low blood pressures and underestimates high blood pressures. The diverging pattern of measurements at clinical extremes and the wide limits of agreement suggest that non-invasive MAP measurements should be interpreted with caution, particularly when making clinical decisions in patients with very low or very high blood pressures. This is a critical finding as underestimation of blood pressure by NIBP, particularly at higher blood pressures, could lead to inadequate management of hypertension in sICH (and less so in TBI) patients. Conversely, overestimation at lower ranges might result in delayed recognition and treatment of hypotension, further exacerbating secondary brain injury [23]. Although the deleterious effects of hypertension in TBI are less well established [22], early hypotension has been associated with a poor outcome [21].
Error grid analysis demonstrated that the above mentioned under- and overestimations are of clinical relevance. For TBI patients, 6.1% of the MAP measurements fell in the ‘moderate to dangerous risk’ categories, indicating that measurement inaccuracies are of a magnitude that they could result in harmful treatment decisions, or omission of treatment. For sICH patients this number was even as high as 12.5%. Although treatment decisions are generally made based on physiological trends rather than single measurements, our findings suggest that at any given time-point non-invasive monitoring may potentially alter therapeutic decision-making in a significant proportion of patients, thereby affecting the efficacy of pre-hospital neuroprotective care [11, 24,25,26]. Furthermore, over 90% of the TBI patients in our study required PHEA. Peri-PHEA NIBP measurement accuracy is important not only to deliver a timely, safe and effective emergency anaesthetic [27], but also to establish post-PHEA maintenance of anaesthesia through infusion therapies and appropriate vasopressor dosages [28, 29].
Pre-hospital IBP monitoring is critical for maintaining haemodynamic optimisation in TBI or sICH patients receiving PHEA, especially during extended transport intervals. Despite minor scene delays from arterial line insertion [24, 30], the substantial clinical benefits strongly justify this approach before PHEA administration. In our cohort, 153/198 (77.3%) patients received arterial access pre-PHEA, demonstrating concordance with established in-hospital anaesthesia guidelines [31, 32]. Pre-hospital invasive monitoring does present challenges, including vascular complications and damping phenomena that can compromise treatment decisions [33]. Clinicians must carefully weigh these limitations against the significant advantages of precise blood pressure management when caring for neurologically injured patients in the pre-hospital environment. Further, arterial line monitoring may facilitate serial blood gas analysis allowing for precise ventilation management [34], more accurate titration of vasopressors, titration of post PHEA anaesthesia and arterial waveform analysis. These extend the benefit of arterial monitoring beyond haemodynamic optimisation, to broader neuroprotection strategies.
Our study has several limitations. First, the retrospective design limits our ability to control for potential confounding factors and (selection) bias. For example, the average measurements per patient was 9 (IQR 3–15, range: 1–29) with a ‘pair’ defined anywhere within a 60 second time-interval. Although the frequency largely represents clinical practice dependant upon clinical need and transport duration, it does mean that patients with more measurements would contribute disproportionately to the analysis. Second, for 100,006 IBP data pairs, no concurrent NIBP measurement was available. Although this could largely be attributed to the higher sampling frequency, in some patients NIBP was likely discontinued when IBP measurements were available, with potential for systematic bias. Third, in some patients NIBP may not have been recorded as they were in a critically hypotensive- or hypertensive state. Although excluding these patients may have resulted in a selection bias, it is an additional argument to support pre-hospital invasive blood pressure monitoring. Fourth, although we attempted to exclude artefactual measurements, the dynamic pre-hospital environment may introduce measurement errors that are difficult to eliminate in retrospect. Despite following established protocols for arterial line setup and maintenance, we cannot retrospectively quantify the prevalence or impact of subtle waveform damping that may have influenced IBP measurements. While grossly abnormal waveforms would have been identified and corrected by clinicians, minor damping effects could still have affected absolute pressure values without being detected. This limitation is inherent to retrospective analysis of arterial pressure monitoring data. Future prospective studies should include systematic documentation of damping coefficients to further characterise measurement inaccuracy. Fifth, a limitation specific to EGA pertains to the geometric shape and zonal placement within the error grid. Marginally different arterial pressure measurements may be assigned to different zones termed ‘zone skipping’ [18]. Further, error grids are derived from a survey among twenty-five experts in the field of anaesthesiology and intensive care medicine lacking response validation, therefore the zones are based on perceived and not real-time clinical action and may only predict the clinical relevance of discrepancy in this cohort. Finally, the single-centre design limits generalisability to other systems with varying resources, geographical challenges and standard operating procedures. In our cohort, median GCS was 6, and 94.7% received PHEA, therefore the merit of IBP may prove less in patients with greater haemodynamic stability.
Conclusion
Pre-hospital non-invasive blood pressure to guide haemodynamic optimisation of patients with TBI or sICH is hampered by clinically relevant measurement inaccuracies in a significant proportion of patients. We find that pre-hospital invasive blood pressure monitoring has the potential to improve early haemodynamic optimisation, especially when hypo- or hypertension is present, supporting one component of tailored neuroprotection in the hyperacute phase.
Data availability
No datasets were generated or analysed during the current study.
Abbreviations
- AAGBI:
-
Association of Anaesthetists of Great Britain and Ireland
- CI:
-
Confidence Interval
- DBP:
-
Diastolic Blood Pressure
- EGA:
-
Error Grid Analysis
- EMS:
-
Emergency Medical Services
- GCS:
-
Glasgow Coma Scale
- HEMS:
-
Helicopter Emergency Medical Services
- HRA:
-
Health Research Authority
- IBP:
-
Intra-arterial Blood Pressure
- IQR:
-
Interquartile Range
- LA:
-
Limits of Agreement
- MAP:
-
Mean Arterial Pressure
- MTC:
-
Major Trauma Centre
- NIBP:
-
Non-invasive Blood Pressure
- NIHR:
-
National Institute of Health and Care Research
- iDBP:
-
Invasive Diastolic Blood Pressure
- iSBP:
-
Invasive Systolic Blood Pressure
- KSS:
-
Air Ambulance Charity Kent Surrey Sussex
- OR:
-
Odds Ratio
- RCoA:
-
Royal College of Anaesthetists
- SBP:
-
Systolic Blood Pressure
- sICH:
-
Spontaneous Intracranial Haemorrhage
- STROBE:
-
Strengthening the Reporting of Observational Studies in Epidemiology
- TBI:
-
Traumatic Brain Injury
References
Rangel-Castilla L, Gasco J, Nauta HJW, Okonkwo DO, Robertson CS. Cerebral pressure autoregulation in traumatic brain injury. Neurosurg Focus. 2008;25(4):E7.
Davis DP, Olvera D, Selde W, Wilmas J, Stuhlmiller D. Bolus Vasopressor Use for Air Medical Rapid Sequence Intubation: The Vasopressor Intravenous Push to Enhance Resuscitation Trial. Air Medical Journal. 2023;42(1):36–41.
O’Horo JC, Maki DG, Krupp AE, Safdar N. Arterial catheters as a source of bloodstream infection: a systematic review and meta-analysis. Crit Care Med. 2014;42(6):1334–9.
Perera Y, Raitt J, Poole K, Metcalfe D, Lewinsohn A. Non-invasive versus arterial pressure monitoring in the pre-hospital critical care environment: a paired comparison of concurrently recorded measurements. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine. 2024;32(1):77.
Sende J, Jabre P, Leroux B, Penet C, Lecarpentier E, Khalid M, et al. Invasive arterial blood pressure monitoring in an out-of-hospital setting: an observational study. Emerg Med J. 2009;26(3):210–2.
Franschman G, Verburg N, Brens-Heldens V, Andriessen TMJC, Van der Naalt J, Peerdeman SM, et al. Effects of physician-based emergency medical service dispatch in severe traumatic brain injury on prehospital run time. Injury. 2012;43(11):1838–42.
Hansen LH, Ettrup-Christensen A, Bülow K. Feasibility of continuous noninvasive arterial pressure monitoring in a prehospital setting, measurements during emergency transfer. European Journal of Emergency Medicine. 2019;26(5):334–9.
Wildner G, Pauker N, Archan S, Gemes G, Rigaud M, Pocivalnik M, et al. Arterial line in prehospital emergency settings – A feasibility study in four physician-staffed emergency medical systems. Resuscitation. 2011;82(9):1198–201.
Butterfield ED, Price J, Bonsano M, Lachowycz K, Starr Z, Edmunds C, et al. Prehospital invasive arterial blood pressure monitoring in critically ill patients attended by a UK helicopter emergency medical service- a retrospective observational review of practice. Scand J Trauma Resusc Emerg Med. 2024;32(1):20.
Eichlseder M, Labenbacher S, Pichler A, Eichinger M, Kuenzer T, Zoidl P, et al. Is time to first CT scan in patients with isolated severe traumatic brain injury prolonged when prehospital arterial cannulation is performed? A retrospective non-inferiority study. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine. 2024;32(1):81.
Irisawa T, Karamooz M, Stolz U, Murphy RA, McDannold R, Venuti M, et al. Compliance with prehospital traumatic brain injury guidelines is poor with longer prehospital treatment duration. Resuscitation. 2014;1(85):S14.
Lewis PS. Oscillometric measurement of blood pressure: a simplified explanation. A technical note on behalf of the British and Irish Hypertension Society. J Hum Hypertens. 2019;33(5):349–51.
Chan V, Pole JD, Keightley M, Mann RE, Colantonio A. Children and youth with non-traumatic brain injury: a population based perspective. BMC Neurology. 2016;16(1):110.
Juri T, Suehiro K, Kimura A, Mukai A, Tanaka K, Yamada T, et al. Impact of non-invasive continuous blood pressure monitoring on maternal hypotension during cesarean delivery: a randomized-controlled study. J Anesth. 2018;32(6):822–30.
von Elm E, Altman DG, Egger M, Pocock SJ, Gøtzsche PC, Vandenbroucke JP, et al. The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studies. Epidemiology. 2007;18(6):800–4.
Myles PS, Cui J. Using the Bland-Altman method to measure agreement with repeated measures. Br J Anaesth. 2007;99(3):309–11.
Haghayegh S, Kang HA, Khoshnevis S, Smolensky MH, Diller KR. A comprehensive guideline for Bland-Altman and intra class correlation calculations to properly compare two methods of measurement and interpret findings. Physiol Meas. 2020;41(5): 055012.
Saugel B, Grothe O, Nicklas JY. Error Grid Analysis for Arterial Pressure Method Comparison Studies. Anesth Analg. 2018;126(4):1177–85.
ANSI AAMI ST79 2017. Available from: https://www.aami.org/standards/featured-standards/ansi-aami-st79
Berger A. Oscillatory Blood Pressure Monitoring Devices. BMJ. 2001;323(7318):919.
Zafar SN, Millham FH, Chang Y, Fikry K, Alam HB, King DR, et al. Presenting blood pressure in traumatic brain injury: a bimodal distribution of death. J Trauma. 2011;71(5):1179–84.
Krishnamoorthy V, Chaikittisilpa N, Kiatchai T, Vavilala M. Hypertension After Severe Traumatic Brain Injury: Friend or Foe? J Neurosurg Anesthesiol. 2017;29(4):382–7.
Dhillon NK, Huang R, Mason R, Melo N, Margulies DR, Ley EJ, et al. Vasopressors in traumatic brain injury: Quantifying their effect on mortality. The American Journal of Surgery. 2020;220(6):1498–502.
Dinsmore J. Traumatic brain injury: an evidence-based review of management. Continuing Education in Anaesthesia, Critical Care and Pain. 2013;13(6):189–95.
Feldman O, Samuel N, Kvatinsky N, Idelman R, Diamand R, Shavit I. Endotracheal intubation of COVID-19 patients by paramedics using a box barrier: A randomized crossover manikin study. PLOS ONE. 2021;16(3): e0248383.
Patrick C, Ward B, Anderson J, Fioretti J, Rogers Keene K, Oubre C, et al. Prehospital Efficacy and Adverse Events Associated with Bolus Dose Epinephrine in Hypotensive Patients During Ground-Based EMS Transport. Prehosp Disaster Med. 2020;35(5):495–500.
Kouz K, Bergholz A, Timmermann LM, Brockmann L, Flick M, Hoppe P, et al. The Relation Between Mean Arterial Pressure and Cardiac Index in Major Abdominal Surgery Patients: A Prospective Observational Cohort Study. Anesth Analg. 2022;134(2):322–9.
Turner J, Bourn S, Raitt J, Ley E, O’Meara M. Pre-hospital emergency anaesthesia in the United Kingdom: an observational cohort study. British Journal of Anaesthesia. 2020;124(5):579–84.
Mohindru J, Griggs JE, de Coverly R, Lyon RM, ter Avest E. Dispatch of a helicopter emergency medicine service to patients with a sudden, unexplained loss of consciousness of medical origin. BMC Emerg Med. 2020;25(20):92.
Fok PT, Teubner D, Purdell-Lewis J, Pearce A. Predictors of Prehospital On-Scene Time in an Australian Emergency Retrieval Service. Prehospital and Dis Med. 2019;34(3):317–21.
Lockey DJ, Crewdson K, Davies G, Jenkins B, Klein J, Laird C, et al. AAGBI: Safer pre-hospital anaesthesia 2017. Anaesthesia. 2017;72(3):379–90.
Guidelines for the Provision of Emergency Anaesthesia Services 2024 | The Royal College of Anaesthetists [Internet]. [cited 2024 Sep 29]. Available from: https://www.rcoa.ac.uk/gpas/chapter-5
Romagnoli S, Ricci Z, Quattrone D, Tofani L, Tujjar O, Villa G, et al. Accuracy of invasive arterial pressure monitoring in cardiovascular patients: an observational study. Crit Care. 2014;18(6):644.
Knapp J, Doppmann P, Huber M, Meuli L, Albrecht R, Sollid S, et al. Pre-hospital endotracheal intubation in severe traumatic brain injury: ventilation targets and mortality—a retrospective analysis of 308 patients. Scand J Trauma, Resuscitation Emerg Med. 2023;31(1):46.
Acknowledgements
We are grateful for the support of all clinicians and operational support colleagues at Air Ambulance Charity Kent Surrey Sussex for providing clinical context and operational experience to support the implementation of invasive arterial blood pressure monitoring at our service. We would also like to thank the clinicians, research team and senior management at the South-East Coast Ambulance Trust for our ongoing partnership. A special thank you to Professor Oliver Grothe at the Institutions of Operations Research, Karlsruhe Institute of Technology, Karlsruhe, Germany who created and provided correspondence on the MATLAB computer program allowing complementary Error Grid Analysis.
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Author information
Authors and Affiliations
Consortia
Contributions
All authors fulfilled the CRediT taxonomy for authorship. Conceptualisation: JG/EtA/RG/RL. Methodology: JG/EtA. Data curation: JG/SC. Formal analysis: JG/EtA. Methodology: JG/RG/AW/SHB/MD/KH/JB. Writing – review and editing: JG/EtA/RL. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
All data were routinely collected and met Health Research Authority (HRA, UK) criteria as service evaluation. Research Ethics Committee approval was not required. The project was approved by the KSS Research and Innovation Committee and registered with the University of Surrey. Study conduct was in accordance with Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) guidelines.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
13049_2025_1393_MOESM1_ESM.docx
Additional file 1. Univariate and multivariate analysis of variables associated with pairwise disagreement in mean arterial pressure (>10%), systolic blood pressure (>20%) and diastolic blood pressure (>20%) in in patients with suspected TBI (n=159) and sICH (n=50).
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Griggs, J.E., Clarke, S., Greenhalgh, R. et al. Diagnostic accuracy of pre-hospital invasive arterial blood pressure monitoring for haemodynamic management in traumatic brain injury and spontaneous intracranial haemorrhage. Scand J Trauma Resusc Emerg Med 33, 89 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13049-025-01393-4
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13049-025-01393-4